![Knick PORTAVO 904X PH User Manual Download Page 29](http://html.mh-extra.com/html/knick/portavo-904x-ph/portavo-904x-ph_user-manual_1980164029.webp)
29
Data Logger
Data Logger Menu
Logger view
Select using arrow keys, confirm by pressing
set
.
CONT
Select start address and start the data logger
START
Deletes all entries and starts the data logger at start
address 0001
DEL
Deletes all entries
SET
Selects and configures logger type (see table
below)
set
Data Logger
Configuring the Data Logger
Prerequisite: The data logger is stopped (press
meas
).
Measurement
Measured value is maintained
Logger: CONT blinks
Logger: START blinks
Logger: DEL blinks
Logger: SET blinks
Logger: Current logger type
blinks
Select desired logger type using
: DIFF,
INT, DIFF+INT or SHOT.
set
Select the appropriate parameters using
and confirm each selection by press-
ing
set
. When configuration is finished, CONT blinks. You can start the data logger
by selecting START or CONT (see page 31).
Summary of Contents for PORTAVO 904X PH
Page 2: ......
Page 50: ...50 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 51: ...51 pH 02 Knick CaliMat Nominal values in bold C pH Buffer Tables ...
Page 52: ...52 pH 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated Buffer Tables ...
Page 53: ...53 pH 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added Buffer Tables ...
Page 55: ...55 pH 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH Buffer Tables ...
Page 56: ...56 pH 07 WTW Technical Buffers Nominal values in bold C pH Buffer Tables ...
Page 57: ...57 pH 08 Hamilton Nominal values in bold C pH Buffer Tables ...
Page 58: ...58 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 59: ...59 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...
Page 66: ...66 V Value added features 8 Viewing recorded data 32 Viewing the logger data 32 Index ...
Page 67: ......