Cal SOP Calibration Method
The calibration method must be configured using the Paraly SW 112 PC software.
Here, you specify which buffers are to be used in which sequence. You can combine
buffer solutions from different buffer sets. Please note that the minimum distance
allowed between two buffer solutions is Δ 2 pH units.
SOP calibration allows you to:
• select up to three calibration points and three buffer sets.
• add a verification buffer.
• specify a maximum deviation (0 ... 0.5 pH units) for the verification buffer as delta
pH.
Sensor Verification
The Paraly SW 112 PC software allows a sensor to be assigned to the device. See the
Paraly SW 112 PC software user manual.
Setup / Cal / Logger Code
Access codes can be set on the meter or using the Paraly SW 112 PC software;
see page 50.
Configuration: SETUP CODE
Calibration: CAL CODE
Data logger: LOGGER CODE
Without entry of an access code, the data logger will only display logger data (
RCL
).
Temperature Calibration
(also separately available as Option 002 TEMP.CAL)
Option 002 TEMP.CAL (Temperature Calibration)
For Memosens sensors, you can perform a 1-point calibration of the internal
temperature detector. See the Calibration chapter for a description.
Option 001 SOP (Standard Operating Procedure)
Summary of Contents for Portavo 904X COND
Page 2: ......
Page 67: ...67 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 68: ...68 pH Buffer Tables 02 Knick CaliMat Nominal values in bold C pH ...
Page 69: ...69 pH Buffer Tables 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated ...
Page 70: ...70 pH Buffer Tables 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added ...
Page 72: ...72 pH Buffer Tables 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH ...
Page 73: ...73 pH Buffer Tables 07 WTW Technical Buffers Nominal values in bold C pH ...
Page 74: ...74 pH Buffer Tables 08 Hamilton Nominal values in bold C pH ...
Page 75: ...75 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 76: ...76 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...