33
Oxy
Oxygen Calibration
Measurement
cal
CAL
AIR CAL
Place sensor in air and wait until the measured
values have stabilized.
Measured value
Temperature
xxx rH blinks
Use
to set the correct relative humidity
value.
cal
xxx nA blinks
Hourglass blinks
Calibration is performed.
Please wait.
CAL DATA
Date
SLOPE xxx nA
Slope calibration data is displayed.
The meter then automatically returns to measuring mode.
AIR CAL Calibration
(Calibrating the slope in air)
The calibration method is selected in the configuration menu.
Note:
To abort calibration, you can press
meas
at any time.