17
Calibration
Measurement
cal
CAL
CELL CONST.
The conductivity will be shown in the display and
can be compared with a reference solution (tem-
perature-corrected).
Value blinks
Use
to set the cell constant value.
cal
Calibration is performed. Display: CELL CONSTANT.
The meter then automatically returns to measuring mode.
CELL CONST Calibration
(Calibration by entry of cell constant)
The calibration method is selected in the configuration menu.
Measurement
cal
CAL
COND
Dip sensor in solution.
Value blinks
Use
to set the temperature-corrected conduc-
tivity value.
Note
: Here, the meter does not perform
a temperature compensation!
cal
Calibration is performed. Display: CELL CONSTANT.
The meter then automatically returns to measuring mode.
COND Calibration
(Calibration by entry of conductivity)
The calibration method is selected in the configuration menu.