![Knick PORTAVO 904 PH User Manual Download Page 36](http://html.mh-extra.com/html/knick/portavo-904-ph/portavo-904-ph_user-manual_1980161036.webp)
36
Selecting the temperature calibration (TEMP. OFFSET)
In measuring mode, press the set
key.
1) Select CAL (calibration) and confirm by pressing set.
2) Select the TEMP. OFFSET calibration mode and confirm by pressing set.
Performing the temperature calibration (TEMP. OFFSET)
In measuring mode, press the cal key.
Press cal once more to activate the function:
Temperature value currently
measured by the sensor
Use the
st
keys to enter
the reference value.
Indication of currently
adjusted offset value.
Press cal to save the reference value.
Option 002 Temperature Calibration