Option 001 SOP (Standard Operating Procedure) meets specific requirements of the
pharmaceutical and biotechnological industries. It comprises:
1. SOP Cal
In the Paraly SW 112 software, you specify which buffers are to be used in which
sequence. You can combine buffer solutions from different buffer sets.
Please note that the minimum distance allowed between two buffer solutions is
Δ 2 pH.
SOP calibration allows you to:
• use up to 3 buffers for adjustment
• use a 4th point for verification (“verification buffer”)
• specify a maximum deviation from the verification buffer
• use buffers from different buffer sets, including a “user buffer”.
2. Sensor Verification
The Paraly SW 112 software allows a sensor to be assigned to the device.
Connecting another sensor generates an error message (ERR21).
3. Conf / Cal Code
You can specify access codes for configuration and calibration on the device or via
Paraly SW 112 software. On the device, the codes are entered in the configuration
menu: SETUP CODE (access code for configuration) and CAL CODE (access code for
calibration) – see page 17. When accessing the configuration (Setup) or calibration
(Cal) menu, you will be prompted to enter these codes.
4. Temperature Calibration
(also separately available as Option 002 TEMP.CAL)
You can perform a 1-point calibration of the temperature detector included in the
sensor, see following page.
Option 001 SOP