17
Calibration
Measurement
cal
CAL
ZERO CAL
xxx
nA
PRESS CAL blinks
Place sensor in oxygen-free medium (e.g.,
nitrogen 5.0) and wait until the measured
values have stabilized.
cal
CAL DATA
Calibration is performed.
Please wait.
Date
ZERO P. xxx nA
Zero calibration data is displayed.
Date
SLOPE xxx nA
Slope calibration data is displayed.
The meter then automatically returns to measuring mode.
ZERO CAL Calibration
(Zero calibration with oxygen-free medium)
The calibration method is selected in the configuration menu.
Note: To abort calibration, you can press meas at any time.
Calibration