Calibrating
Measurement
CAL
CELL CONST.
Calibration method, the number of calibration
points and the buffer set have been selected in
the configuration menu.
Value blinks
Use
st
to select the value for the
cell constant.
Calibration is performed. Automatic return to measuring mode.
CELL CONST Calibration
(Calibration by entry of cell constant)
The calibration method is selected in the configuration menu.
Measurement
CAL
COND
Dip sensor in solution.
Value blinks
Use
st
to adjust the temperature-corrected
conductivity value. NOTICE: Here, the meter
does not perform a temperature compensation!
Calibration is performed. Automatic return to measuring mode.
COND Calibration
(Calibration by entry of cell conductivity)
The calibration method is selected in the configuration menu.