15
Conductivity Configuration
Prior to measurement, a configuration should be performed to match
the connected sensor and the desired measurement performance.
Furthermore, you can select the suitable calibration method.
The following table gives you an overview.
Factory settings are shown in bold print.
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing set.
Configuring
Measurement
“Setup” display
Select using arrow keys, confirm by pressing set.
Display
Cond | Conc %
|
SAL g/kg | TDS mg/l | °C
MOHM cm
OFF | On
Cond Unit
mS/cm | S/m
TDS Factor
0.0 … 1.0 (if display = TDS)
TC
*)
OFF | LINEAR | NLF | NACL | HCL | NH3 | NAOH
(if display = Cond)
TC LINEAR
0.0 … 20.0 %/K | 2.1 %/K (if TC = LINEAR)
REF. Temp.
0 … 100 °C | 25 °C
(
32 … 212 °F | 77 °F)
(if TC = LINEAR)
Conc. Table
When display has been set to “Conc %”: -01- ... -10-
(for concentration ranges, see page 39)
CAL
CELL CONST. | COND | 0.01 MOL KCL | 0.1 MOL KCL |
INST. FACTOR** | ZERO POINT** | FREE CAL
Auto OFF
OFF | 0.1h | 1h | 6h | 12h
Temp. Unit
°C | °F
Time Format
24h | 12h
Date Format
dd.mm.yy | mm.dd.yy
TAN TEMP CAL
(TAN input required, option; see page 31)
Default
NO | YES (reset to factory settings)
*) Temperature compensation **) For inductive cond. measurement only