MemoTrans
24
9.1.1 Changing the Parameters
The device functions according to the plug & play principle of Memosens technology. However, to
ensure that the settings of the previous sensor type do not get lost, the new sensor type must be
selected in the software.
Path:
Guidance
Measurement parameter
Function
Options
Info
Measurement parameter
Selection
• pH, ORP, pH/ORP
• Dissolved oxygen
• Conductivity
Select the parameters supported by the device.
9.2 Adapting the Measuring Device to the Process Conditions
Medium Compensation (in the Process) for Oxygen
Path:
Application
Sensor
Extended setup
Function
Options
Info
Medium pressure
Selection
• Process pressure
• Air pressure
• Altitude
Specify altitude (‑300
…
4000 m),
process pressure (500
…
9999 hPa),
or air pressure (500
…
1200 hPa) of measuring point.
Salinity
0
…
40 g/kg
Factory setting
0 g/kg
The influence of salt content on oxygen measurement
is compensated with this function.
Example: sea water measurement as per Copenhagen
Standard (30 g/kg).