61
Configuring the Time/Date
Time and Date
Control of the calibration and cleaning cycles is based on the time and date of the integrated real-time
clock.
In measuring mode the time is shown in the lower display. When using digital sensors, the calibration
data is written in the sensor head. In addition, the logbook entries (cf Diagnostics) are provided with a
time stamp.
Note:
There is no automatic switchover from winter to summer time!
Be sure to manually adjust the time!
Sensor Verification (TAG, GROUP)
When Memosens sensors are calibrated in the lab, it is often useful and sometimes even mandatory that
these sensors will be operated again at the same measuring points or at a defined group of measuring
points. To ensure this, you can save the respective measuring point (TAG) or group of measuring points
(GROUP) in the sensor. TAG and GROUP can be specified by the calibration tool or automatically entered
by the transmitter. When connecting an MS sensor to the transmitter, it can be checked if the sensor con-
tains the correct TAG or belongs to the correct GROUP. If not, a message will be generated and Sensoface
gets “sad”. The “sad” Sensoface icon can also be signaled by a 22 mA error current. Sensor verification can
be switched on in the Configuration in two steps as TAG and GROUP if required.
When no measuring point or group of measuring points is saved in the sensor, e.g., when using a new
sensor, Stratos enters its own TAG and GROUP. When sensor verification is switched off, Stratos always
enters its own measuring point and group. A possibly existing TAG/GROUP will be overwritten.
Menu item
Action
Choices
TAG
of measuring point
In the lower display line you can enter a desig-
nation for the measuring point (TAG) and for a
group of measuring points (GROUP) if applica-
ble. Up to 32 digits are possible.
By pressing meas (repeatedly) in the measur-
ing mode you can view the tag number.
Select character using
keys,
select next digit using
keys.
Press enter to confirm.
A...Z, 0...9, – + <
> ? / @
The first 10 char-
acters are seen in
the display with-
out scrolling.
GROUP
of measuring points
Select number using
keys,
select next digit using
keys.
Confirm by pressing enter