![Knick A405B User Manual Download Page 38](http://html.mh-extra.com/html/knick/a405b/a405b_user-manual_1980154038.webp)
38
Current range
Process variable
LIN/biLIN/LOG output
Current start
Current end
Bilinear: Vertex X
Bilinear: Vertex Y
Time averaging filter
Output current during error message
Output current for
Sensoface message
Output current during HOLD
Output current for HOLD FIX
Output Current: Curve
Example: Current Output 1
1 Press menu key.
2 Select CONF using
,
press enter.
3 Select OUT1 menu using
keys,
press enter.
4 All items of this menu group are indicated by the
“OT1:“ code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) by pressing enter.
5 Exit: Press meas key until the [meas] mode indicator is
displayed.
Configuring the Current Output