32
Device Type: Cond / CondI
1 Press menu key.
2 Select CONF using
,
press enter.
3 Select SENSOR menu using
keys,
press enter.
All items of this menu group are indicated by the
“SNS:” code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) by pressing enter.
4 Exit: Press meas key until the [meas] mode indicator is
displayed.
Select measuring mode
Select measuring range
Concentration determination
Temperature unit
CHECK TAG
CHECK GROUP
enter
Configuration
The device type is automatically selected upon first start-up. In the SERVICE menu you can change the
device type. Afterwards, you must select the corresponding calibration mode in the CONF menu.