Technical data
Vacuum Pump System SC/SEM/SH/SR950
10
Translation of original Operating Instructions, English, KNF 125549-125552 07/18
Other parameters
Vacuum connection and gas
outlet
For tube ID 10 mm
Ventilation connection
For tube ID 4 mm
Weight [kg]:
- SC 950 (including hand
terminal)
- SEM 950
- SH 950
- SR 950
14
13.5
13
12.5
Dimensions W x H x D [mm]:
SC/SEM/SH 950:
SR 950:
246 x 487 x 313
246 x 487 x 264
Maximum permissible ambient
relative humidity
80 % for temperatures up to
31°C, decreasing linearly to
50 % at 40°C
Maximum altitude of
installation
[m above sea level]
2000
Only SC/SEM/SH 950:
Operating parameters of coolant (high-performance condenser)
Max. permissible pressure
[bar g]
3
Permissible temperature
- 15 °C to + 20 °C
Coolant connections on high-
performance condenser
For tube ID 8 mm
Coolant-wetted surface [cm
2
]
Min. 470
Electrical data vacuum pump system
Voltage** [V]
100-240 V +/- 10 %
Frequency [Hz]
50-60 Hz
Max. current consumption [A]
1.7
Max. watt consumption [W]
150
Protection class
IP20
Fuse vacuum pump system [A] 2 x T2.5
Drive motor fusing
Electronic overcurrent protection
Tab. 2 (part 2)
** Automatic voltage selection
The pumps of the vacuum pump systems are supplied by a
universal power supply with integrated overload protection.
They are protected against overheating by a temperature
sensor on the motor board and equipped with overcurrent
protection.
If one of these safety functions is triggered, the pump will be
shut down and must be manually reset, as follows:
Separate pump from the mains.
Remove the cause(s) of the fault before restarting.