Operation
20
7.6
Preparing a measurement
Note:
The device should be switched on at least 5 minutes before use.
To check the stirring wire, press the <VIB> button.
Stab injury
Risk of injury from stirring wire during function test performance.
Perform the function test only with the sample tube attached.
Note:
If you want to print out the measurement results, you may connect
a printer via RS-232 connector.
The result of each measurement will be printed automatically in the fol-
lowing format:
Measured 0400 mOsmol
23.03.15 08:24:35
After a calibration run, you get the following print:
Calibrated 1234 mOsmol
23.03.15 07:36:17
If a measurement is canceled due to an error, the screen displays the
cor responding error message. If a printer is connected, it prints the error
message, without the specification about the type of error.
(NO) FREEZE
23.03.15 09:06:04
7.6.1
Measurement preparation process
The device is switched on and calibrated.
1.
Pipet 50–150 µl sample or calibration solution into a clean, dry sample
tube.
Practical Tip:
The sample volume can be freely chosen in the mentioned
range. It is important that you use exactly the same amount for calibration
and measurement.
Note:
Please consider, the higher the volume, the better the preci sion of
the measurement.