26
Operation
KNAUER
10D User Manual V7670, Version 2.3
Control
Choose between analog or digital ports to control the device
accordingly.
Ports
LAN DHCP: Control with LAN, IP address is obtained auto-
matically (recommended)
RS-232 9600: Serial port for cables longer than 2 m
RS-232 115200: Serial port for cables with up to 2 m length
ANALOG: Control with applied voltage
Prerequisite
Where applicable: a physical connection has been estab-
lished.
Status display is active.
Activating/Deactivating the Lamp
The status display shows not only absorption and wavelength,
but also indicates the status of the lamp.
HEA: deuterium lamp lights up
OFF: Deuterium lamp is inactive.
ON: Deuterium lamp is active.
Prerequisite
Status display is active.
Steps
Figure
Procedure
1. Keep left arrow key
pressed. Press right
arrow key until the cor-
rect display appears.
2. Let go of left arrow key.
3. Using both arrow keys,
choose a port.
Fig. 14
Display port
Result
When setting is finished, the status display appears.
Control
LAN DHCP
Process
Figure
Procedure
1. Keep left arrow key
pressed. Press right
arrow key until the cor-
rect display appears.
2. Let go of left arrow key.
3. Using both arrow keys,
activate or deactivate
the deuterium lamp.
Fig. 15
Display Lamp
Result
When setting is finished, the status display appears.
LAMP D2
OFF