AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 31
7.3
GLP
The following GLP data of the detector can be found in the software.
GLP data
Unit
Explanation
Device information
Serial number
FRAYYWWXXXXX
Firmware version
Current firmware version of device
Operating time
h
Operating hours after production
Date of installation
Date of manufacture
Last maintenance
Light source
Operating time
h
Running hours after light source
installation
Date of installation
Light source number
Indicates how often the light
source has been changed since
manufacture.
Validation data
Last measured span
µRIU
Last measured span
date
Leak sensor
Serial number
Serial number leak sensor
Firmware version
Current firmware version of leak
sensor
7.4
Switching on the detector
The recommended warm-up time for the detector is 60 minutes. This rec-
ommendation is applicable after turning the detector on as well as after
powering up from standby.
The rear side must be accessible.
Detector has been switched off.
Burns
An overconcentration of organic solvents may lead to explosions.
Check the tighness of all fi ttings before using organic solvents.
Device defect
Changes of the environmental temperature cause condensation inside
the device.
Allow device to acclimate for 3 h before connecting to power supply
and taking into operation.
Prerequisites