AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 15
Procedure
Figure
3.
Insert the capillary as far as
possible into the IN port
4
or
OUT port
5
at the front of the
device.
4.
Tighten the fitting finger tight.
4
5
Check if all fittings are tight.
6.2.2 AZURA® Detector RID 2.1L HighFlow
At the front side, both connections are designed for 1/4'' UNF flat bottom
fittings.
Open-end wrench
Component defect
Damage of the flow cell by high backpressure.
Install the backpressure regulator from the supplied kit.
IN port
Procedure
Figure
1.
Push the capillary
1
through
the fitting
2
.
2.
Slide the lock ring
3
over the
tubing. Note that the tapered
end of the lock ring has to
point to the seal ring
4
.
3.
Attach the seal ring
4
.
1
2
4
3
2
1
3
2
4
4.
Insert the capillary as far as
possible into the IN port
5
at
the front of the device.
5.
Tighten the fitting finger tight.
Next steps
Tools
5