Technical Data
Removing a Leak
Prerequisite
If liquid enters the inside of the device, switch off the device. If
this is not the case, it is not necessary to switch off the device.
Auxiliary material
Cloth for drying the leak sensor
Next steps
Bring the device into operation again.
Technical Data
P 6.1L Isocratic
P 6.1L Binary
Procedure
Process
1. Remove the leak.
2. Dry the leak tray.
3. Acknowledge the system error via Control Unit, Mobile
Control, or the standby button (only in standalone mode).
Setup
Pump type
Isocratic analytical HPLC pump
Variants of the
pump head
5 ml/min stainless steel
10 ml/min stainless steel
50 ml/min stainless steel
10 ml/min ceramic
50 ml/min ceramic
Delay volume
60 μl (for 10 ml/min)
Weight
Weight
11.5 kg
Setup
Pump type
Binary analytical HPLC pump with degas-
ser
Variants of the
pump head
5 ml/min stainless steel
10 ml/min stainless steel
10 ml/min ceramic
10 ml/min stainless steel
for normal phase applications
Degasser
2 channels
Solvent Selection
valve
2 x 2 channels
Weight
Weight
14.1 kg
Gradient formation
Gradient forma-
tion
High-pressure binary mixing