If you leave the correction factor at 0.0 %, the temperature compensation
is deactivated and the raw data is displayed (regardless of the set referen-
ce temperature).
The monitor has been switched on.
Process
Display
1.
Press both arrow keys.
2.
Hold down the left arrow key and use the right
arrow key to scroll until the correct display
appears.
3.
Let go of the arrow keys and press the left or
right arrow key to select a value.
Enter additional settings or start the analysis.
5.6 Adjusting the calibration function
The calibration points of the instrument are preset at the factory. The ins-
trument was adjusted to 86.6 mS/cm at 25 °C using a 1 M NaCl solution.
Automatic calibration is not possible.
If you require a higher accuracy of the measured conductivity value in a
special range, the calibration function can be adjusted at four freely selec-
table, independent points as follows:
The monitor has been switched on.
Note:
The values to be entered for „
Factor #“ must increase with increa-
sing #.
Process
Display
1.
Flush the flow cell with a solution of known con-
ductivity (actual point).
2.
Set one of the factors to the target point.
Press both arrow keys.
Hold down the left arrow key and use the
right arrow key to scroll until the correct dis-
play appears.
Let go of the arrow keys and press the left or
right arrow key to select a value.
Wait until the status display shows.
Prerequisites
Next steps
Calibration
Prerequisites