AZURA® Assistant ASM 2.2L Instructions V6803
18
Product information
The AZURA® Compact HPLC is based on the multifunctional AZURA®
Assistant ASM 2.2L. The assistant includes a UV / VIS detector with a vari-
able wavelength and a compact pump with pressure sensor. The manual
injection valve is attached to the side of the system.
5.8
Exchange of the device modules
Hot surface
The detector module can become warm during operation.
Allow the detector module to cool for 20 minutes before removing
it from the assistant.
Component defect
When changing the device modules if the assistant is switched-on ,
damage to the board is possible.
Switch off the assistant before changing device modules.
Process
Figure
1. Remove the front cover
1
.
2. Loosen the four loss-resistant
screws
2
crosswise using a
phillips screwdriver (size 1) .
Remove module
1
2