5.2
Performance features
The assistant can be controlled with a chromatography data system (Open-
LAB®, ClarityChrom® and PurityChrom®) or with an optional touch display
(Mobile Control). There are LAN and analogue interfaces available. Thus,
the assistant can be integrated into almost any LC system.
The Mobile Control and supported software products (restricted in
PurityChrom®) display or read GLP data, such as the operating hours of the
device or light sources. For a detailed description of viewing or reading
the GLP data, see the appropriate instructions for the software products
(
https://www.knauer.net/en/Support/User-manuals/Software
).
The modular structure makes it possible to adapt the assistant to individual
requirements. The plug-in modules can be easily released and replaced by
other devic modules (see chapter “5.8 Exchange of the device modules”
on page 18).
The AZURA® Neo electronic platform features:
New microprocessor for faster device performance
Communication interfaces: IP dual stack with switch (for connecting
AZURA® devices among each other) and LAN stack function plus USB
service interface (internal USB to RS-232). Both LAN ports (1 and 2) can
be used as interface or as switch.
Industrial standard 4–20 mA analog input (replacing 0–10 V input on
the previous electronic platform).
No external display support.
5.3
Views
5.3.1
Front view
Legend
1
Device name and
serial numbers
2
UV detector
3
Valve drive
4
Pressure sensor
5
Pump head
6
Module name and
serial number
Fig. 1
ASM 2.2L front view
Control
GLP data
Modular design
AZURA® Neo
6
1
Left Modul
Middle module
Right module
2
3
4
5