AZURA® Assistant ASM 2.2L Instructions V6803
Maintenance and care 41
12.2 System information
Using the Mobile Control or the software you can read out the operating
hours of the assistant.
12.3 Cleaning and caring for the device
All smooth surfaces of the device can be cleaned with a mild, commercially
available cleaning solution, or with isopropanol.
12.4 Maintenance tasks for users
For maintenance, care and the recommended intervals, please refer to the
operating instructions of the respective device modules.
12.5 Preparing the assistant for storage or
transportation
Pay attention that all hoses and capillaries have been emptied or filled with
flushing solution (e. g. isopropanol) before storage. To prevent algae
formation, do not use pure water.
We recommend the same ambient conditions (temperature range and
humidity) for storage as during operation (see ’Technical data’ on p. 44).
Flow cell has been filled with isopropanol.
The assistant is switched off.
Open-end wrench
Operating hours
Device defect
Intruding liquids can cause damage to the device.
Place solvent bottles next to the device or in a solvent tray.
Moisten the cleaning cloth only slightly.
Prerequisites
Tools
Life-threatening injuries
Health danger if getting in contact with toxic, infectious or radio-active
substances.
Before disposing of the device or sending it away for repair, you are
required to decontaminate the device in a technically correct manner.