AZURA® Assistant ASM 2.2L Instructions V6803
34 Operation
Process
Figure
Procedure
1. Switch on the device at the
power switch
1
on the rear
side.
2. Wait until the device has com-
pleted the self-test.
3. If the self-test has been success-
fully completed, the LED
2
on
the right lights up green.
Fig. 16
Power switch on
the rear side of the device
The firmware is successfully initialized if shortly after switching on the
device all three LEDs lights lights up red, green and blue for approx. 1
second. If the test fails an error message will be displayed. If the error
occurs several times in succession, contact KNAUER technical support.
9.4
Activate standby
The flow has been switched off.
Note:
Malfunctioning system after repeated standby possible. After
repeatedly using the standby, switch off the power switch and back on
again, to reset the data storage.
Procedure
Figure
1. Press the standby switch
2
for
5 seconds.
2. Wait if the LED
1
shows a blue
light.
1 2
If you were successful, the right LED lights up blue.
To end standby, press the switch again.
9.5
Switching off assistant
The assistant is designed for the use of different solvents. In case the
assistant has not been used for several weeks, solvent residues may cause
damage. If you plan to take the assistant out of operation for a long period
of time, fill the capillary with alcohol (eg. isopropanol) before switching it
off.
The back must be accessible.
The assistant is switched on.
The capillaries have been flushed.
1
Result
Prerequisites
Result
Next steps
Prerequisites