Unpacking and setup
7. Unpacking and setup
7.1 Location requirements
Note:
Proper operation is only guaranteed if you adhere to the specifica-
tions for the ambient conditions and the place of use.
Before you specify the location, read the technical data (see p. 44). There
you will find all the important information about power connection,
ambient conditions and humidity.
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Set up the device in such a way that it is protected against exposure
to direct sunlight.
Leave room for air circulation: See paragraph „space requirements“.
Set up the assistant on an even surface.
Protect the assistant from direct sunlight.
Set up the assistant in a place protected from draft (air conditioning).
Protect the assistant from strong draft.
Avoid placing the assistant next to machines that cause ground vibra-
tion.
Avoid vibration.
Keep the assistant away from high frequency sources. High-frequency
sources may compromise measuring values.
If you are located in an earthquake area, use the bore holes.
The bore holes are located on both right and left side panels.
Minimum distance of 5 cm with a device on one side.
Minimum distance of 10 cm with devices on both sides.
Minimum distance of 15 cm at the back for the fan.
7.2
Unpacking and setup
Note:
The leak sensor may malfunction it the device is placed on an
inclined surface. Use a level to check that the device is in an horizontal
position.
Store all packing material. Retain included packing list carefully for repeat
orders.
Devi
ce defect
The center of gravity of the assistant is located at the front of the
device.
The device could fall down and be damaged.
Lift the device only centrally on the side of the housing.
General requirements
Space requirements