Maintenance and care

Maintenance and care

Proper maintenance of your HPLC device will ensure successful analyses 
and reproducible results. Please check the user manuals of the micro 
devices to find all information on maintenance (see “Device overview” on 
page 5)
.

Maintenance contract

The following maintenance work on the device may only be performed by 
KNAUER or a company authorized by KNAUER and is covered by a sepa-
rate maintenance contract:

Opening the device. 

Removing the hood or the side panels. 

Required system information

Serial number

The serial number can be found with the Mobile Control (

GLP

device

) as 

well as on the rear side of the device.

Year of manufacture

The year of manufacture forms part of the serial number of the module, 
example: CDA103252525. The letters designate the device type; the year 
of manufacture is stated as year and calendar week. In the example, the 
module was manufactured in calendar week 32 of the year 2010. The last 
five digits are the identification number for the interface card (

Interface: 

Serial number

).

Cleaning and caring for the device

Device defect

Intruding liquids can cause damage to the device.

Place solvent bottles next to the device or in a solvent tray.

Moisten the cleaning cloth only slightly.

All smooth surfaces of the device can be cleaned with a mild, commer-
cially available cleaning solution, or with isopropanol.

Maintenance tasks for users

The following maintenance tasks are recommended by KNAUER and may 
be performed by the user. The procedures are described in the according 
user manuals.

V

Valves have different 
types in 'column switch' 
mode

Correct the settings. Both valves have to be 
of the same type

Valve left: failure 
detected!

Left valve is stiff.

Drive is defective.

Valve middle: failure 
detected!

Middle valve is stiff

Drive is defective.

Valve right: failure 
detected!

Right valve is stiff

Drive is defective.

W

Wrong valve type

Correct the settings.

System message

Solution

Summary of Contents for Azura Assistant ASM 2.1L

Page 1: ...Assistant ASM 2 1L User Manual HPLC...

Page 2: ...AUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 D 14163 Berlin Phone 49 30 809727 0 Fax 49 30 8015010 Internet www knauer net E Mail info knauer net Version information Article number V6815 Version...

Page 3: ...Safety for users 11 Signal words 12 Decontamination 12 Symbols and signs 12 Unpacking and setup 13 Contacting the Technical Support 13 Location requirements 13 Unpacking 13 Connectors on the rear sid...

Page 4: ...ce contract 27 Required system information 27 Cleaning and caring for the device 27 Maintenance tasks for users 27 Technical data 28 Communication 28 General 28 Repeat orders 28 Legal information 30 T...

Page 5: ...s Environmental analyses Eluents Small quantities of substances such as additives modifiers or salts can influence the durability of the materials For further reference on chemical compatibility pleas...

Page 6: ...take a look into the UVD 2 1S user manual V6820 http www knauer net fileadmin user_u pload produkte files Dokumente detectors azura manuals V6820_A ZURA_UVD_2 1S_User Manual_EN pdf Flow cells Flow cel...

Page 7: ...h 6 ports and 3 channels In case of multiposition valves the central port does not count For further information take a look into the V 2 1S user manual V6860 http www knauer net fileadmin user_upload...

Page 8: ...ood or similar extraction device KNAUER Valves Ports Connection Material Pressure stability 2 position valve 6 1 16 1 8 Stainless steel 300 bar 2 position valve 6 1 16 Stainless steel 400 bar 2 positi...

Page 9: ...ight of detector UVD 2 1S because the fittings of the detector flow cell cannot be mounted An assistant with only degasser integrated is not allowed Position of modules and IP port connections The pos...

Page 10: ...Devices ClarityChrom OpenLab Mobile Control V4 Two pumps independent yes no no Fraction valve 1 1 no Injection module1 1 An injection module is a combination of one pump and one 6 port 2 position valv...

Page 11: ...ts tools and solvents made or recommended by KNAUER Good Laboratory Practice GLP Accident prevention regulations published by the accident insurance companies for laboratory work Filtration of substan...

Page 12: ...occur Minor injuries can occur Device defects can occur Decontamination Contamination of devices with toxic infectious or radioactive substances poses a hazard for all persons during operation repair...

Page 13: ...t 10 cm if further devices are set up on both sides At least 15 cm on the rear panel for the fan Leave the power plug on the rear of the device accessible to be able to disconnect the device from the...

Page 14: ...ou notice any damage contact the Technical Support 6 Place the device in its site of operation and remove protective foil Next steps Store packaging and keep the included packing list for repeat order...

Page 15: ...t least 1000 ms START IN TTL input Low active Secure switching threshold at least 1000 ms After receiving a signal short circuit to ground from an external device the device starts If controlled with...

Page 16: ...w on page 5 PEEK fittings Component defect Damage to the ports caused by strongly tightened fittings Use 5 Nm torque for stainless steel fittings Use 0 5 Nm torque for PEEK fittings ASM 2 1L as Compac...

Page 17: ...nt cover Process Process Figure 1 Carefully push the funnel 1 into the center opening of the capillary guide 2 Fig 7 Funnel and capillary guide 2 Push the long ending of the first nozzle 4 into the ho...

Page 18: ...their IP address Prerequisite In Windows power saving hibernation standby and screen saver must be deactived In case you use an USB to COM box the option Allow the computer to turn off ths device to...

Page 19: ...e can be used to remotely control the system Integrating the LAN into a Company Network A network administrator can integrate the LAN into your company net work In this case you use the WAN port of th...

Page 20: ...trol is a program for a computer or tablet To operate via Mobile Control the computer or tablet needs to be connected to a WLAN router You will find a detailed description in the Mobile Control user m...

Page 21: ...odules are at the desired location The power plugs of the modules are plugged in The LAN connections between the modules and router are connected The LAN cable is connected to the workstation and rout...

Page 22: ...ation Qualifica tion is a standardized document and free of charge It is not part of the delivery please contact the Technical Support in case of request The Operation Qualification includes the follo...

Page 23: ...r is integrated into a company network pull out the patch cable from the WAN port Can the devices communicate with the computer even though the router is disconnected from the company network 5 Turn o...

Page 24: ...ves When the pump head seals are defec tive solvent enters the piston backflush ing Inform the Technical Support of KNAUER Exchange the pump head Pressure and flow rate variations Pump without pressur...

Page 25: ...ed offautomatically because Pmax was exceeded Exchange the filter system Pressure increase Replace cartridge Filter system inserted wrong Exchange the filter system Problem Solution System message Sol...

Page 26: ...t Switch the device off and then on If the leak sensor is still not present contact the Technical Support of KNAUER Leak was detected Switch off the device Remove the leak and start the device afterwa...

Page 27: ...endar week In the example the module was manufactured in calendar week 32 of the year 2010 The last five digits are the identification number for the interface card Interface Serial number Cleaning an...

Page 28: ...ts and accessories can be found online www knauer net Module Measures 2000 operating hours Detector Check light intensity of the D2 lamp Inspect the flow cell assembly Clean the flow cell Replace the...

Page 29: ...ec tion valve pH flow cell and 2 small prepacked col umns A9854 1 Mounting bracket AZURA L for axial compressible columns with 20 mm ID A9853 3 Mounting bracket AZURA L for Hypershear mixing chamber A...

Page 30: ...nd transport damage In the event of device malfunctions directly contact the manufacturer KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Germany Phone 49 30 809727 111 Telefax 49 30...

Page 31: ...uld be as small as possible Degasser Degasser module for fluids GLP Good Laboratory Practice quality assurance for labo ratories Gradient Time dependent composition of solvent mobile phase on low pres...

Page 32: ...ASM 2 1L user manual V6815 Abbreviations and terminology 32 Solvents Mobile phase eluent or carrier for liquid chromato graphy Term Explanations...

Page 33: ...M 2 1L 29 DG 2 1S 8 P 2 1S P 4 1S 6 UVD 2 1S 6 V 2 1S 7 Mobile Control 20 29 mounting bracket 29 O operating environment 13 OQ 22 P packing list 14 PEEK 16 port LAN 19 power disconnect 13 plug 13 powe...

Page 34: ...ASM 2 1L user manual V6815 Index 34 V valve 7 W warranty 30...

Page 35: ...R manuals online www knauer net knowledge KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Phone Fax E Mail Internet 49 30 809727 0 49 30 8015010 info knauer net www knauer net KNAUER...

Reviews: