![KLS Martin MCO 25 plus Instructions For Use Manual Download Page 78](http://html1.mh-extra.com/html/kls-martin/mco-25-plus/mco-25-plus_instructions-for-use-manual_1979006078.webp)
78
Revision 5.5
6.10.6
Overlap
Using the parameter
OVERLAP
you can adjust the effect of the super pulse line scan.
A stronger overlap results in a higher power density and a stronger effect.
Effect of the overlap parameter used with line scanning
LIGHT (0%, no overlap)
MEDIUM (50%, half of the area overlaps)
STRONG (80%, large part of the area overlaps)
6.11
Memory Functions
6.11.1
Storing of Parameters
NOTICE
When switching from Time Mode to Frequency Mode and back, the stored parameters are
lost. The names of the storage locations can be carried over.
Using the memory keys
M1…M5
the laser parameter settings shown on the
display are stored and loaded. You can store
up to 5 different settings. The parameters
remain stored even when the laser device is
switched off.
Saving the settings
To save the current settings, press a memory
key (
M1…M5
) and hold it down for about 5 s.
Fig. 6-14:
Storing the setting CW,
15 W, pilot brightness 55%