Revision 5.5
63
6.7.2
Scanner Mode Parameters
Setting range for maximum power
from 10–25 W in steps of 1 W.
Figure size
Use the
SELECT
key to select the field
FIGURE SIZE
. Now increase or decrease the value
using the arrow keys.
Setting ranges
The size range of the scan shapes depends on the scanner handpiece connected to the
scanner.
Shape
127 mm
200 mm
Triangle
2–8 mm
3–13 mm
Square
2–8 mm
3–13 mm
Rectangle
4–8 mm
6–13 mm
Hexagon
2–8 mm
3–13 mm
Circle
2–8 mm
3–13 mm
Thick ring
6–8 mm
9–13 mm
Thin ring
4–8 mm
6–13 mm
Straight line
1–6 mm
1–9 mm
Curved line
1–6 mm
1–9 mm
Strongly curved line
1–6 mm
1–9 mm
Tab. 6-4:
Figure size of scan shapes for scanner handpieces
NOTICE
The energy density is the important parameter for the effects on the tissue.
When the size of the scan shape is changed, the set energy density is not. This means that
the effects on the tissue remain constant.
Settings for scan speed
low, middle, high
The exposure times of the laser per location resulting from the speeds are below the thermal
tissue relaxation time of 0.7 ms. The scanning speed is doubled when changed from low to
middle and again from middle to high.
NOTICE
If the energy density is constant, the thermal strain on the tissue is reduced with increased
scan speed.