34
Revision 5.5
CAUTION
Potential danger to the patient from interlock deactivation!
Opening the operating room door may otherwise result in
unforeseeable interruption of the treatment and thus in a
dangerous situation for the patient!
If interlock deactivation is not desired, the interlock connector
must be bridged with an internal jumper.
Function:
•
Opening the operating room door
does not
cause the
device to switch into STANDBY mode.
•
The laser treatment can be continued without interruption.
5.6
Preparing the MCO 25 plus Laser for Attaching the
Mirror Joint Arm
•
Unscrew protection cap from the thread (1).
•
Remove gray cap from the opening of the 5-mm hexagon spanner (2).
•
Turn the clamping (4) that is located under the cap to the left using the 5-mm hexagon
spanner (3).
—
The clamping for the mirror joint arm is turned to the side.
—
The arm can be mounted.
Fig. 5-2:
Clamping for installation of the mirror joint arm
(opened by turning to the left, closed by turning to the right)
1 Thread for mirror joint arm (here without cover)
2 Opening for 5-mm hexagon spanner (gray cap)
3 5-mm hexagon spanner
4 Clamping