![KLS Martin maXium smart C Instructions For Use Manual Download Page 119](http://html1.mh-extra.com/html/kls-martin/maxium-smart-c/maxium-smart-c_instructions-for-use-manual_1979004119.webp)
maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
119
16
Guidelines and Manufacturer’s Declaration on Electromagnetic
Compatibility (EMC)
Guidance and manufacturer’s declaration – Electromagnetic interference emission
The device maxium® smart C with or without maxium® smart Beam is designed for use in an electromagnetic
environment as specified below. The user must ensure that it is operated in such an environment.
Immunity tests
Compliance level
Electromagnetic environment – Guidelines
HF emissions according to CISPR 11
Group 2
The device must emit electromagnetic energy in
order to fulfill its intended function. Nearby
electronic units may be affected.
HF emissions according to CISPR 11
Class B
The requirements of the class are complied with
only in operation readiness without activation of
HF current!
Emission of harmonics according to
IEC 61000-3-2
Class A
The device is suitable for use in any facilities,
including residential areas and such as are
directly connected to the normal public supply
network for residential buildings.
Emissions of voltage fluctuations/flicker
according to IEC 61000-3-3
Complies