Suspension Arm Systems Independant
Revision 1
73
7.7.3
Connecting the Earthing Cables
Fig. 7-31:
Connecting the earthing cables (Single and Single/Duo)
For reasons of simplification, the illustrations show the interface plate
without extension
arm or cables.
•
Observe the safety notices, see section 7.7.1 “Safety Notes”, page 70.
•
Cut all green-yellow earthing cables to 2.5 mm² and 10 mm² in length.
•
Feed the earthing cables through the tension relief and connect them to the series terminal
4 mm² or 10 mm² on the earthing terminal unit on the interface plate
.
Ensure that the earthing cables are safely in place in the tension relief :
•
All earthing cables must sit securely in the tension relief .
7.7.4
Connecting the Power Cables
Fig. 7-32:
Connecting the power cables (Single and Single/Duo)
For reasons of simplification, the illustrations show the interface plate
without extension
arm or cables.
•
Observe the safety notices, see section 7.7.1 “Safety Notes”, page 70.