Electrostatic Field Meter EFM 022  

(V 5.3)

 

 

System Prof. Dr.-Ing. Hans  Kleinwächter 

Summary of Contents for EFM 022

Page 1: ...MANUAL Electrostatic Field Meter EFM 022 V 5 3 System Prof Dr Ing Hans Kleinw chter...

Page 2: ...le 6 Application Areas 6 Button Features 6 Initiation 6 Hold 6 Measurement ranges 7 E Field Meter Measurement 7 Max Value 7 Changing the measurement distance 8 Spacers 8 Measurement with mounted volta...

Page 3: ...l collecting electrons is charged negative There are different ways to avoid or to discharge electrostatic charges But to find an effective and reasonable solution first of all the emergence amount an...

Page 4: ...er DAC Resolution 10Bit Battery 9V alkaline or NiMH battery Operating time Approx 10h with alkaline battery Measuring range Voltmeter Distance Measuring range Max resolution 1cm 0 10kV 1V 2cm 0 20kV 2...

Page 5: ...5 Caption Caption 1 3 2 4 5 6 7 8 1 Cover cap 5 Spacer 2 Modulator system 6 Zero point trimmer 3 Button function on 7 Alphanumeric LCD display 4 Grounding jack 8 Battery case on the back...

Page 6: ...briefly to change to a new distance or to CPS mode wait approx 2s to select the displayed function Aus Press function on 2 times in quick succession Initiation By pressing the function on button on t...

Page 7: ...e select the E Field Meter Mode In this Mode it s possible to select the following ranges 20KV m 200kV m or 1MV m To change the ranges press the function on key briefly The range changes as follows In...

Page 8: ...rect voltage field strength Over the selected dis tance the field strength is converted to the charge on the object displayed value V field strength V m x distance m Example With a distance of 10cm an...

Page 9: ...ention Use only Rechargeable 9V NiMH Batteries or Alkaline Batteries Batteries must be charged out of the device in a suitable battery charger Please pay attention to the manufacturers instructions Us...

Page 10: ...evice Finally the battery compartment should be closed again Waste disposal Batteries and accumulators are not allowed in the household waste Every consumer must by law dispose of all batteries and ac...

Page 11: ...rd Certificate of calibration Optionally accessories CPS 022 Charge Plate Set To test air ionization systems by measuring the discharge time EFM 023 It has the same function as the EFM 022 with an add...

Page 12: ...plosive areas If high charges are possible the electrostatic field meter must be grounded An adequate distance must be kept Flashovers on the modulator system must be avoided The use of the device in...