![KLA Tencor P-16+ User Manual Download Page 66](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647066.webp)
Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-18
KLA-Tencor Confidential
0142530-000 AB
3/13/09
If the Stylus chosen is small enough to detect the features of interest in the scan,
then a sampling rate and scan speed should be optimized to accurately record the
level of detail required from the scan.
Figure 3.14
Data Collection When Using a
Large Radius
Stylus
Stylus Path A (large radius stylus) -
Low Frequency Data Collection
Low and High Frequency Data Traces
are nearly the same since a large radius
stylus was used.
Stylus Path A (large radius stylus) -
High Frequency Data Collection
Low Frequency Data Trace
High Frequency Data Trace
Sample Surface