![KLA Tencor P-16+ User Manual Download Page 405](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647405.webp)
0142530-000 AB
KLA-Tencor Confidential
14-7
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Configuration - Stage Configuration
Theta Soft Home Position
14
The Soft Home position is set at manufacturing and should not require adjustment.
This procedure should only be attempted by a KLA-Tencor trained technician.
Teach Lowest Elevator Position
14
Introduction
14
The Lowest Elevator Position sets the vertical motion range of the stage. Using this
feature, a limit (
Z coordinate
) can be set for the elevator so that the measurement
head cannot descend past the level of the sample surface.
Correctly teaching the Lowest Elevator Position protects the measurement head when
the Proximity Sensor (which is used to switch from
Elevator Focus Speed
to
Elevator
Slow Focus Speed
) is not being used.
Figure 14.7
Stage Configuration Parameters
CAUTION:
It is very important to reset the correct
Lowest Elevator
Position
after a precision locator is installed. The stylus can be
damaged if the stage remains configured to the original setting.