I. Basics
I. Basics
I. Basics
I. Basics
Scope of Chiller supply:
-KSC215-L-…
-Installing Instruction inside the switch cabinet
-Manual inside the switch cabinet
-Data transfere cable 50m (164feet) in a box near the pump
-Short data transfere cable in a box near the pump
-Grounding in a box near the pump
-Overvoltage protection in a box near the pump
-Spare parts in a box near the pump
-Brass fittings for the connection Chiller inlet/outlet and piping (on the pump)
-Stainless steel fittings for the connetion piping and IFP (on the pump)
Carefully read the operating instructions located in the control cabinet before beginning
installation.
Check the equipment for damage on arrival and report any defects immediately.
Claims submitted later cannot be honored.
Please observe the following notes and warnings.
1.
Removing the operating instructions from the cooling block voids the warranty!
2.
The volume of the water circulation system of the cooling block is approx. 17 liters
(~4.5 gal) Please consider this by filling the water system with water ethylene
glycol mixture.
3.
Always operate the water circulation system at a volume of 35-38% ethylene
glycol.
This is regardless of the ambient temperature.
Non-compliance voids the warranty.
Use of automobile anti-freeze and propylene glycol is prohibited.
Fill with clean water (potable water quality) with Ethylene-glycol at the rate of min.
35% to max. 38%.
Use ethylene-glycol of only one manufacturer.
(e.g.: DOWTHERM SR1-Dow Chemical, Safeflow EG-Clariant)
Do not mix two ore more different manufacturer for one water-ethylene-glycol
circuit.
Open front panel and fill to a pressure of min.1,5 bar via the feed cock when the
pump is off.