KITO Harrington DHLT Calibration Procedure Download Page 1

Dynamic Hoist Load Tester (DHLT)

Calibration Procedure

 
 

H

ARRINGTON 

H

OISTS

,

 

I

NC

.

 

C

ONFIDENTIAL

 

1 of 4 

General Use

EDOC0430

Rev. 01  10-5-2016

 

P

URPOSE

 

To provide instructions for calibration of the DHLT. 
S

COPE

 

(1)

 

Includes the DHLT.  Excludes the SHLT (Static Hoist Load Tester). 

(2)

 

Includes only calibration.  Excludes setting the Pressure Switch and setting the Limit Switch. 

(3)

 

Calibration based on using Harrington 5T Electric Chain Hoist (either NER050L or ER050L). 

I

NSTRUCTIONS

 

1.

 

Collect Calibration Data 

(a)

 

Assemble 5 Ton Electric Chain Hoist and properly calibrated Force Measuring Device (Load 
Cell or Dynamometer) to the 5 Ton link in the test stand (see Fig. 1). 

(b)

 

Connect power to test stand control panel. 

(c)

 

Connect pendant to test stand control panel. 

(d)

 

Connect cable between test stand control panel and hoist. 

(e)

 

Connect power to the hoist. 

(f)

 

Ensure the Accumulator Discharge Valve located between the accumulator and the reservoir 
is closed (handle is perpendicular to the hoses). 

(g)

 

Press the On/Off pushbutton on the test stand control panel to turn on the DHLT.  The DHLT 
pump should begin operating and maintain system pressure between approx 1700-2300 psi. 
For DHLT units shipped after 10/2010, the pump should begin operating and maintaining 
system pressure between approx. 1700 – 2450 psi.  

(h)

 

Turn on the Dynamometer; assure that it is freely suspended from the Test Hoist with no 

tension on it (other than the weight of the attached lower shackle). Press its zeroing button 

and ensure the read-out indicates zero load.   

Note: Do not press the zeroing button any time after testing starts. 

(i)

 

Use the Test Hoist to level the lower beam assembly; this minimizes any effects associated 

with slant-lifting. 

(j)

 

Follow the calibration procedure for both the 5 ton and 10 ton links as follows.   Using the 
Pressure Control Valve (PCV), slowly adjust the cylinder pressure to increase the system 
pressure.  An increase of system pressure can be verified by the Cylinder Pressure Gauge 
(CPG). See the last page of this document for a data recording chart.  

Note: Small movements in the PCV correspond to large changes in the Dynamometer 
reading, particularly on the 10 Ton link; avoid large or rapid adjustments to the PCV as 
these can lead to pressure oscillations.   

Note: During the following procedure, if you go above the target cylinder pressure while 
adjusting “up”, reduce the cylinder pressure back down below the target value, and then 
raise it up to the target value again.  Similarly, if you go below the target cylinder 
pressure while adjusting “down”, raise the cylinder pressure back above the target value, 
and then lower it down to the target value again. 

Reviews: