22
9.
Warranty / Recycling
Warranty
:
The breast pump is guaranteed for 3 years from the date of purchase against any manufacturing defect,
except case of misuse.
For after sales service under warranty, please contact DTF MEDICAL or your local distributor.
The technical documentation is available on request.
Recycling
:
The device must be disposed of in accordance with local regulation on the disposal of electrical
appliances. So you contribute to the human health and environmental protection. Product on the
market since August the 13
th
2005.
10.
Technical specifications
Device class (breast pump)
IIa
Device class in electrical safety (breast pump)
II
Type of applied parts
BF
Protection level
IP 30
Expected shelf life of accessories
1 year
Expected shelf life of autoclavable accessories
50 cycles
Expected shelf life of the breast pump
5 years
Collection container capacity
100ml
Weight
4 kg
Dimensions
L360 x l250 x h180mm
Voltage
Available in 230V – 50Hz; 220V – 50Hz or 220V – 60Hz
Consumption
100VA
Fuses
T 0,5A H 250V
Suction values
Variable from -50 to -300hPa
Cycles
40 to 125 cycles/minutes
Sound level
<57dBa
Running mode
Continuous
Power supply
Alternative current 230V – 50Hz
Ambient temperature (T) - For transport and storage
-25°C < T < +70°C
Ambient temperature (T) - When in functioning
+5°C < T < +40°C
Relative humidity (H)
For transport, storage and functioning
15% < H < 93%
Atmospheric pressure (P)
For transport, storage and functioning
700 hPa < P < 1060 hPa
Essential performance features:
•
Provide adequate suction so that mothers can express their milk.
•
Control the rate of suction to best meet the baby’s sucking rhythm.
Applied part:
Breast shield.
Identification:
Each breast pump has a unit identification (reference and number) which indicates, among other things, the
exact date of manufacture and the serial number of the breast pump. The first two digits show the year of
manufacture (e.g.
68
: 1950 +
68
= 2018) and the next three digits indicate the day of the year of manufacture
(e.g. 285 = 12
th
October). The last two digits indicate the serial order number in the series. In addition, each
accessory is identified by a lot number mentioned. For more information, refer to the instructions for use of
the KIT EXPRESSION.
Summary of Contents for FISIO PRO
Page 61: ...61 21 93 42 CEE 14 1993 II 3 2 5 0 BF Off I On 13 2005 BPA...
Page 65: ...65 6 V Hz 3000 KIT EXPRESSION 43 5 10 KOLOR 7 DTF MEDICAL DTF MEDICAL...
Page 66: ...66 KOLOR KIT EXPRESSION KIT EXPRESSION c 1 9 1 120 I O d 3 5 a www dtf fr KOLORYOU b FISIO PRO...
Page 68: ...68 4 4 2 S 76 L 90 KITETT araroKoK KOLORYOUR 2 S L 3 a b KOLOR 20 10 10...