background image

Inner Ear Thermometer

 IET2

User Instructions

The ear thermometer is a device capable of achieving infra-red temperature measurement when placed in
the auditory canal of a subject. It is a safe means of measuring human body temperature through the ear.

Specification:

  Range of displayed temperature : 

34

~44

(93.2

~111.2 )

      Display L

(L

) when the temperature is under 34

(93.2

)

      Display H

(H

) when the temperature is over 44

(111.2 )

  

Operating ambient:

      Temperature: 16

~35

(60.8

~95.0

)

      Humidity: 

80%rh

  

Storage and transport condition:

      

Temperature: -10

 to 55

(14

 to 131

)

      Humidity: 30%rh to 90%rh

  

Display resolution temperature range:

 0.1

(0.1

)

  

Accuracy:

 

0.2

 (from 35.5

 to 42.0

)

                      

0.4

 (from 96.0

 to 107.6

)

  

Power consumption:

 30  mW

  

Battery: 

two 1.5V alkalescence batteries (AAA)

  

Battery life:

 4000 takes or 6 months with 3-min.usege per day

  

Dimension:

 134mm by 37mm  by 67mm

  

Net weight: 

74g

  

Beeper sign: 

on/off, measuring finish and low voltage warning, etc.

  

Self-testing sequence: 

Press the '

on/recall'

 button to turn on the thermometer and all of symbol 

                                             (See Fig A) should be displayed on the LCD in one second. 

  

Calibration: 

This thermometer is calibrated at the time of manufacture. If the thermometer is used 
according to the instructions, periodic re-calibration is not required.

  

Manufacturing Batch code:

see the label within the battery compartment.

Probe membrane

Disposable probe cover

 Probe cap

On/recall button

Batterry cover

Scan button

LCD

  

Display:

 liquid crystal display, 4 digits

      temperature value: display the maximum temperature

                                   in measuring process

      temperature unit: centigrade or fahrenheit
      display of memory: last ten memories
      low battery  warning: the LCD display           and then a beep sound is heard

Please read the instructions carefully before using.

                                                              Cautions

1. Condition the thermometer at room temperature for 30 minutes before taking a reading.

2. Use a new probe cover with each and every measurement to prevent inaccurate readings.

3. Please keep your ear canal clear, if not, the measuring result shall be inaccurate.

4. The

5. Do not scratch the probe membrane as this will affect the accuracy of measurements.

6. If the LCD does not change or respond when you press any of the buttons, remove the batteries 
    and then replace to reset.

12. This device must always be kept in a clean, dry area.
13.

      

 

Degree of protection against electric shock is Type BF applied part.

 '     ' symbol on the display indicates the device is ready for use.

14. 

 Battery disposal should be completed in accordance with local regulations.

     :  

 7. Please don't  use the thermometer if your ear canal has become  inflamed.  Do  not  force  the  thermometer
    into the ear canal. If pain occurs, stop using immediately.

8. Do not expose this thermometer to electric shock.
9. Do not expose the thermometer to sunlight or to water.
10. Do not use near strong electromagnetic fields, i.e. Keep it away from any radio systems and mobile phones.

  

11. Do not modify this device without authorization of the manufacturer.

IET2 IB UK 20180525

The standard we adopted is EN12470-5:2003 Clinical thermometers -- Part 5: Performance of infrared
ear thermometers (with maximum device).
This device complies with the requirements of IEC 60601-1,IEC 60601-1-11 and IEC 60601-1-2.

Guidance and manufacturer's declaration - electromagnetic immunity

The IET2 Device is intended for use in the electromagnetic environment specified below. 
The customer or the user of the IET2 Device should assure that it is used in such an environment.

IMMUNITY test

IEC 60601

test level

Compliance

level

Electromagnetic environment - guidance

Radiated RF

IEC 61000-4-3

3 V/m 

80 MHz to 2.5 Ghz

3 A/m

0.01

0.1

1

10

100

0.12

0.38

1.2

3.8

12

0.12

0.38

1.2

3.8

12

0.23

0.73

2.3

7.3

23

Portable and mobile RF communications equipment should
be used no closer to any part of the 

IET2 Device, 

including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter.
Recommended separation distance :

Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation distance 
in meters (m).Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,

a

 should be

less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked 
with the following symbol:

 

d

   1.2  

P

d

   1.2  

P

d

   2.3  

P

d

   2.3  

P

80 MHz to 800 MHz
800 MHz to 2,5 GHz

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted the
oretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
IET2 Device is used exceeds the applicable RF compliance level above, the IET2 Device should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the IET2 Device.
B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

The IET2 Device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the IET2 Device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the IET2 Device as recommended below, according to
the maximum output power of the communications equipment.

Recommended separation distances between

portable and mobile RF communications equipment and the IET2 Device

Rated maximum output

power of transmitter

W

Separation distance according to frequency of transmitter

150 kHz to 80 Mhz

d

   1.2  

P

80 MHz to 800 Mhz

800 MHz to 2,5 Ghz

Reviews: