Response Page Commands
Epi2.0 Response Menu:RESPONSE> ?
Commands:
SCALE: Shows Scale Factor for Each Range and Axis
STATUS: Prints Current Sensor Response Values
SAFE: Disables Sensor System Control
RETURN: Return to Previous Menu
?: Help for info on specific command
Seconds=4421
Ticks=222
Epi2.0 Response Menu:RESPONSE>
The “status” command prompts the system to print out a set of system-related data:
Epi2.0 Response Menu:RESPONSE> status
Sensor FS Range is 4g
X Sensor Module Scale Factor is 5001 mV per Standard g
Y Sensor Module Scale Factor is 4998 mV per Standard g
Z Sensor Module Scale Factor is 4999 mV per Standard g
AUTOZERO is OFF
CAL is OFF
Seconds=4447
Ticks=789
Epi2.0 Response Menu:RESPONSE>
Activity Timeout and Disable
Once enabled, the sensor will automatically be disabled after 3600 seconds of inactivity (no commands
entered), or on a RESET or POWER-ON-RESET event. When disabled, the sensor is returned to the Main
Page. The user would then need to re-enable the system to access other pages. System control is also
disabled with a “safe” command entered on any page.
Seconds and Ticks
The sensor measures and reports elapsed time since its last RESET or Power-ON-RESET event. “Seconds”
are the number of elapsed seconds. “Ticks” are a sub-second measure approximately equal to 1 msec.
Shipping and Handling
While it is called a “strong motion” accelerometer, the EpiSensor2 is still a precision instrument that
should be handled with care. Avoid strong shocks during shipment and installation. It is recommended
that customers utilize the original foam-filled packaging during any shipment and transport.
Contact Kinemetrics
Please contact Kinemetrics with any questions or issues regarding this product.
1-626-795-2220 (Phone)
www.kinemetrics.com
Page 18
August 2015