![Kimo KIGAZ 210 Software Operation Manual Download Page 31](http://html1.mh-extra.com/html/kimo/kigaz-210/kigaz-210_software-operation-manual_1967501031.webp)
14. Perform a measurement of Max CO
14.1. Perform the measurement
➢
Connect a CO probe (available as option) on C1 connection on the top of the analyser or connect the flue gas probe on the
single connector on the bottom of the analyser.
➢
Turn on the analyser.
➢
Go to the
“Measurement”
menu with the keypad then press
OK
.
The analyser performs the autozero: remaining time and a progress bar display on the
screen.
➢
Go to
“Max CO”
with the keypad then press
OK
.
➢
Press
OK
to launch the measurement.
The measurement is launched, the instrument displays the level of CO and the level of
Max CO.
When the measurement duration is finished, the analyser displays the value of CO
maximum during the measurement.
The default measurement duration is
30 s
. It is possible to change this duration before starting
the measurement:
➢
On
“Max CO”
screen, before pressing OK to start the measurement, press the
“Duration”
function key.
“Duration”
screen is displayed.
➢
Press
OK
.
➢
Use the keypad to select the required measurement duration:
30 / 40 / 50 s
or
1 / 2 / 3 / 4 / 5 minutes.
➢
Press
OK
to validate the measurement duration.
➢
Press
Esc
to back to the screen of measurement start of Max CO.
14.2. Link the measurement to a customer
When the measurement has been performed, it is possible to link it to a customer registered in the analyser or to a customer
unregistered in the analyser that must be created.
➢
Press the
“Save”
function button.
The analyser displays the list of recorded customers.
➢
Select a customer with up and down arrows then press
OK
or create a customer (see page 23) then select it.
The analyser displays the list of available boilers.
➢
Select the boiler with up and down arrows then press
OK
or create a boiler (see page 24) then select it.
The analyser backs to the measurement screen.
14.3. Add the measurement to a group
The group function has to be activated (see page 18).
Once the measurement is linked to a customer and a boiler, it is possible to add it in a group in the instrument, or create a group:
The analyser proposes to add the measurement to a group.
➢
Go to
YES
and press
OK
.
➢
Go to the required group with the arrows and press
OK
, or create a group (see page 25).
The analyser backs to the measurement screen.
14.4. Print the results
It is possible to print the results on the ticket:
➢
Press the
“Print”
function button.
Results are printed on the ticket.
See page 17 for the printing configuration.
Perform a measurement of Max CO
31