TOS9200/9201
Basic Operations 3-49
3.8.2
Ending the test
PASS judgement
Example of an AC withstanding voltage test (PASS)
When the timer is ON:
When the test time has elapsed without the occurrence of the FAIL status, the test
ends with a PASS judgement.
When a PASS judgement is made, “PASS” appears at the top right of the LCD, the
PASS LED lights on the indicator, and a buzzer sounds. The PASS judgement is dis-
played for approximately 0.2 seconds (by default). The display time for the PASS
judgement can be set to 0.2 s through 10.0 s or to HOLD. The buzzer action syn-
chronizes with the PASS display time. For the setting procedure, see "3.10 System
Settings".
The measurement results are displayed while “PASS” is displayed.
When “PASS HOLD” is not set to “HOLD,” the tester automatically returns to the
READY status after a PASS judgement is made.
When the timer is OFF:
The test continues until the FAIL conditions are met.
Press the STOP switch to suspend the test. The test does not make a judgement and
“PASS” is not displayed.
Fig.3-18 Timing chart for PASS judgement
(The start voltage is set on the ACW.)
DISP
2.50
kV
11
A
3.0
s
ACW
ACW
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
PASS
NORM
60Hz
START
HV ON
TEST
PASS
READY
100 ms
Rise Time
Test Time
Fall Time
In PASS judgment
Output
voltage
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......