3-26 Basic Operations
TOS9200/9201
Setting the voltage rise time (RISE TIME)
The rise time between the start voltage and the test voltage can be set to 0.1 s
through 200 s (at a resolution of 0.1 s for the 0.1 s to 99.9s range and 1 s for the 100
s to 200s range).
When the cursor is to the right of RISE, settings can be made using the rotary knob.
1.
To move the cursor to the right of RISE, press the F2 key (RISE TM).
(The keys can also be used.)
2.
Using the keys, move the cursor to the target digit.
3.
Using the rotary knob, set the rise time.
The
fi
gure (V/s) in parentheses represents the voltage increase per second calculated
for reference using the set value. If the value cannot be displayed in three digits, “--
-” is displayed instead.
Setting the WAIT TIME
In DC withstanding voltage testing, if a test voltage is applied to a DUT that con-
tains capacitive elements, a large charge current may
fl
ow until charging is com-
pleted. To avoid upper fail judgement by the charge current, a wait time must be
provided from the starts of START VOLTAGE, and upper fail judgement will be
ignored during wait time.
Set the wait time to 0.3 s through 10.0 s (at a resolution of 0.1 s).
When the cursor is located to the right of WAIT, the rotary knob can be used to
make settings.
1.
To move the cursor to the right of WAIT, press the F4 key (WAIT TM).
(The keys also can be used.)
2.
Using the rotary knob, set the wait time.
START V RISE TM
WAIT TM
DCW2
DCW2
START : 10%
RISE
: 0.1s (– – – V/s)
GND :
LOW
WAIT :
0.3s
F1 F2 F3 F4 F5
SHIFT
START V RISE TM
WAIT TM
DCW2
DCW2
START : 10%
RISE
: 0.1s (– – – V/s)
GND :
LOW
WAIT :
0.3s
F1 F2 F3 F4 F5
SHIFT
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......