TOS9200/9201
Basic Operations 3-25
3.5.2
Settings on the DCW2 screen
Setting the start voltage
For a DC withstanding voltage test, set the start voltage as a percentage of the test
voltage, to 0% through 99% (at a resolution of 1%). When the cursor is located to
the right of START, settings can be made using the rotary knob.
1.
To move the cursor to the right of START, press the F1 (START V) key
(the keys can also be used).
2.
Using the rotary knob, set the start voltage.
• When the start voltage is set to 1% or more, approximately 100 ms is required for
the tester to reach the start voltage from 0 V.
Fig.3-7 Start voltage (DCW)
START V RISE TM
WAIT TM
DCW2
DCW2
START : 10%
RISE
: 0.1s (– – – V/s)
GND :
LOW
WAIT :
0.3s
F1 F2 F3 F4 F5
SHIFT
NOTE
RISE TIME
TEST TIME
Time
0
TEST VOLTAGE
START VOLTAGE
Approximately
100 ms