3-18
Basic Operations
TOS9200/9201
3.4.3
Settings on the ACW3 screen
Channel settings for the high-voltage scanner
Make settings with the optional scanner connected.
Each channel can be set to HIGH, LOW, or OPEN.
Use the keys to move the cursor ( ) to a channel, and then use the F1, F2, and
F3 keys to make settings.
If the channel has already been speci
fi
ed, settings can be made using the rotary
knob.
All channels can be opened using the SHIFT + F1 keys, regardless of the location of
the cursor.
• If an unconnected channel is selected, “INVALID CH”
fl
ashes on the LCD.
•
If the test leadwire is not connected to the DUT, do not leave the test lead-
wire connected to the scanner’s output terminal.
The TOS9220 scanner does not have a contact check function to check
connections to the DUT. If the tester is set to the high-voltage (HIGH) level,
a test can be started using a channel with the test leadwire not connected
to the DUT.
• To clearly indicate the relationships between the connected test leadwire and the
channel, af
fi
x the Channel Display Seal (provided for the scanner) to the test lead-
wire.
ACW3
ACW3
F1 F2 F3 F4 F5
SHIFT
HIGH
OPEN
LOW
CH :
HIGH :
LOW :
5 6 7 8
1 1 1
9 0 1 2
1 1 1 1
3 4 5 6
1 2 3 4
CONTACT CHECK : ON
NOTE
ACW3
ACW3
HIGH
OPEN
LOW
CH :
HIGH :
LOW :
5 6 7 8
1 1 1
9 0 1 2
1 1 1 1
3 4 5 6
1 2 3 4
CONTACT CHECK : ON
INVALID CH
WARNING
NOTE
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......