3-12 Basic Operations
TOS9200/9201
Setting the voltage rise time (RISE TIME)
The rise time between the start voltage and the test voltage can be set to 0.1 s
through 200 s (at a resolution of 0.1 s for the 0.1 s to 99.9s range and 1 s for the 100
s to 200s range).
When the cursor is to the right of RISE, settings can be made using the rotary knob.
1.
To move the cursor to the right of RISE, press the F2 key (RISE TM).
(The keys can also be used.)
2.
Using the keys, move the cursor to the target digit.
3.
Using the rotary knob, set the rise time.
The
fi
gure (V/s) in parentheses represents the voltage increase per second calculated
for reference using the preset value.
If the value cannot be displayed in three digits, “---”is displayed instead.
Setting the voltage fall time (FALL TIME)
When the test ends with a PASS judgement, the voltage falls to 0 V during the preset
voltage fall time. This voltage fall time can be set to 0.0 s through 200 s (at a resolu-
tion of 0.1 s for the 0.1 s to 99.9s range and 1 s for the 100 s to 200s range).
When the cursor is to the right of FALL, settings can be made using the rotary knob.
1.
To move the cursor to the right of FALL, press the F3 key (FALL TM).
(The keys can also be used.)
2.
Using the keys, move the cursor to the target digit.
3.
Using the rotary knob, set the fall time.
The
fi
gure (V/s) in parentheses represents the voltage fall per second calculated for
reference using the set value.
If the value cannot be displayed in three digits, “---” is displayed instead.
START V RISE TM FALL TM
ACW2
ACW2
START : 10%
V RANGE : AUTO
RISE
: 0.1s (– – – V/s)
RESPONSE : SLOW
FALL : 0.0s (– – – V/s)
GND
: LOW
F1 F2 F3 F4 F5
SHIFT
START V RISE TM FALL TM
ACW2
ACW2
START : 10%
V RANGE : AUTO
RISE
: 0.1s (– – – V/s)
RESPONSE : SLOW
FALL : 0.0s (– – – V/s)
GND
: LOW
F1 F2 F3 F4 F5
SHIFT
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......