TOS9200/9201
Basic Operations 3-11
3.4.2
Settings on the ACW2 screen
Setting the start voltage
For an AC withstanding voltage test, set the start voltage as a percentage of the test
voltage, to 0% through 99% (at a resolution of 1%). When the cursor is located to
the right of START, settings can be made using the rotary knob.
1.
To move the cursor to the right of START, press the F1 key (START V).
(The keys can also be used.)
2.
Using the rotary knob, set the start voltage.
• When the start voltage is set to 1% or more, approximately 100 ms is required for
the tester to reach the start voltage from 0 V.
Fig.3-2 Start voltage (ACW)
START V RISE TM FALL TM
ACW2
ACW2
START : 10%
V RANGE : AUTO
RISE
: 0.1s (– – – V/s)
RESPONSE : SLOW
FALL : 0.0s (– – – V/s)
GND
: LOW
F1 F2 F3 F4 F5
SHIFT
NOTE
RISE TIME
FALL TIME
TEST TIME
Time
0
TEST VOLTAGE
START VOLTAGE
Approximately
100 ms