2-6 Precautions on Handling
TOS9200/9201
2.7
Daily Checking
To avoid accidents, con
fi
rm at least the following before starting operation:
• The tester is connected to an earth ground.
• The coating of the high-voltage test leadwire is free from cracks,
fi
ssures,
and breakage.
• The high-voltage test leadwire is not broken.
• The tester generats FAIL signal when the ends of the low-voltage test lead-
wire and high-voltage test leadwire are short-circuited.