TOS9200/9201
Setup
1-3
1.2
Precautions for Installation
Be sure to observe the following precautions when installing the tester.
■
Do not use the tester in a f
lammab
le atmosphere
.
T
o pre
v
ent e
xplosion or
fi
re, do not use the tester near alcohol, thinner
, or other
comb
ustible materials, or in an atmosphere containing such v
apors.
■
A
v
oid locations where the tester is e
xposed to high
temperatures or direct sunlight.
Do not locate the tester near a heater or in areas subject to drastic temperature
changes.
Operating temperature range:
+5 °C to +35 °C (+41 °F to +95 °F)
Storage temperature range:
-20 °C to +70 °C (-4 °F to +158 °F)
■
A
v
oid humid en
vir
onments.
Do not locate the tester in a high-humidity en
vironment—near a boiler
, humidi
fi
er
,
or w
ater supply
.
Operating humidity range:
20 % to 80 % RH
(no de
w condensation permitted)
Storage humidity range:
90 % RH or less
(no de
w condensation permitted)
Condensation may occur e
v
en within the operating humidity range. In that case, do
not start using the tester until the location is completely dry
.
■
Do not place the tester in a corr
osive atmosphere
.
Do not install the tester in a corrosi
v
e atmosphere or one containing sulfuric acid
mist or the lik
e.
This may cause corrosion of v
arious conductors and imperfect con
-
tact with connectors, leading to malfunction and f
ailure, or in the w
orst case, a
fi
re.
■
Do not locate the tester in a dusty en
vir
onment.
Dirt and dust in the tester may cause electrical shock or
fi
re.
■
Do not use the tester where ventilation is poor
.
This tester features a forced-air cooling system. Pro
vide suf
fi
cient space for the air
inlet on the lateral side and the air outlet on the rear side to allo
w air to
fl
o
w
.
■
Do not place the tester on a tilted surface or in a location
subject to vibrations.
If placed on a non-le
v
el surf
ace or in a location subject to vibration, the tester may
f
all, resulting in damage and injury
.
■
Do not use the tester in locations aff
ected b
y str
ong ma
gnetic
or electric f
ields.
Operation in a location subject to magnetic or electric
fi
elds may cause the tester to
malfunction, resulting in electrical shock or
fi
re.
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......