TOS9200/9201
Specifications 8-1
Chapter 8
Speci
fi
cations
This chapter describes the electrical and mechanical speci
fi
cations for the tester.
Withstanding Voltage test mode
Item
Speci
fi
cations
TOS9200
TOS9201
Output section
AC
Output-voltage range
0.05 kV to 5.00 kV
Resolution
10 V
Accuracy
±(1.5% of s 20 V) [with no load]
Maximum rated load (*1)
500 VA (5 kV/100 mA)
Maximum rated current
100 mA [output voltage of 0.2 kV or more]
Transformer capacity
500 VA
Output-voltage waveform(*2)
Sine wave
Distortion
2% or less [with no load or pure resistive load
at output voltage of 0.5 kV or more applied]
Frequency
50 Hz/60 Hz
Accuracy
±0.1%
Voltage regulation
±3% or less [maximum rated load
→
no load]
Short-circuit current
200 mA or more, 350 mA or less [at output voltage of 0.5 kV or more]
Type of output
PWM switching
DC
Output-voltage range
––––––––
0.05 kV to 6.00 kV DC
Resolution
––––––––
10 V
Accuracy
––––––––
±(1.5 % of s 20 V)
Maximum rated load (*1)
––––––––
50 W (5 kV/10 mA)
Maximum rated current
––––––––
10 mA
Ripple
No load at 5 kV
––––––––
50 Vp-p Typ.
Maximum rated load
––––––––
150 Vp-p Typ.
Voltage regulation
––––––––
1 % or less [maximum rated load
→
no load]
Short-circuit current
––––––––
40 mA Typ.
Discharge function
––––––––
Forced discharge at the end of test
(discharge resistance: 125 k
Ω
)
Start voltage
The voltage at the start of the test can be set as the start voltage.
Setting range
0% to 99% of the test voltage (resolution of 1%)
Output-voltage monitoring function
If the output voltage exceeds ±(10% of s 50 V), output is cut off
and the protection function activates.
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......