TOS9200/9201
Using Terminals and Connectors 4-1
Chapter 4
Using Terminals and Connectors
This chapter describes the procedures for use of the connectors on the front and rear
panels.
•
In remote control, the tester turns an extremely high voltage on/off using
an external signal. This poses grave potential danger. Provide full safety
measures to prevent the unintentional generation of a high voltage. While
a high voltage is being generated, the DUT, high-voltage test leadwire,
probe, and output terminals never touched. Never conduct remote control
unless suf
fi
cient safety measures have been provided.
• When the START signal of the SIGNAL I/O connector is valid, the START signal
of the SIGNAL I/O has priority over that of the REMOTE terminal.
The STOP signals of the SIGNAL I/O connector and the REMOTE terminal are
accepted equally.
REMOTE terminal on the front panel
The optional remote-control box RC01-TOS/RC02-TOS or high-voltage test probe
HP01A-TOS/HP02A-TOS can be used to start and stop a test.
SIGNAL I/O connector on the rear panel
By inputting a signal to this connector, a test can be started and stopped, and the
panel memory and program memory can be recalled. In addition, using the output
signal from the SIGNAL I/O connector, the status of the tester can be checked.
INTERLOCK connector
Using a signal from an external safety device, the tester can be placed in the
INTERLOCK status.
STATUS OUT connector
This connector is used to connect to a warning light and certain other devices. A
voltage of +24 V is output under the conditions (of the tester) selected on the Sys-
tem screen.
WARNING
NOTE
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......